1 . Modularization . Each module is 4 channels 120V50A . Expandable for 8 channels / 12 channels , etc . in one cabinet .

2 . Parallelled . Up to 4 channels to be parallelled connection , can reach 200A in one module .

3 . Regeneration . The regenerative function is included to greatly reduce power consumption during the discharge test.

4 . Communication; Able to log parameters from BMS, Chamber, I/O signals as a condition for cut-off or protection during testing. Support CANbus, RS485 protocol, SMbus, I²C.

5 . Simulation; Because of high dynamic response speed, it supports the profile dynamic simulation, such as DCIR, HPPC, NEDC, FUDC, and other tests.

The CE-6008n-20V50A-H series battery testing system supports channels to expand . Multi – channel can be chosen . You can add the module according to your real test requirement . Each module has built – in parallel channels . The parallel function aims to increase the max current in real tests , saving the cost to expand the large range . The parallel function also is safety and reliability . Furthermore , withing using powerful BTS8.0 software which independent develop by Neware , you can test the driving simulation test , HPPC test , DCIR test , capacity test , and other tests required high sample data rate , high accuracy , and high response speed .

 

1 . Modularization . Each module is 4 channels 120V50A . Expandable for 8 channels / 12 channels , etc . in one cabinet .

2 . Parallelled . Up to 4 channels to be parallelled connection , can reach 200A in one module .

3 . Regeneration . The regenerative function is included to greatly reduce power consumption during the discharge test.

4 . Communication; Able to log parameters from BMS, Chamber, I/O signals as a condition for cut-off or protection during testing. Support CANbus, RS485 protocol, SMbus, I²C.

5 . Simulation; Because of high dynamic response speed, it supports the profile dynamic simulation, such as DCIR, HPPC, NEDC, FUDC, and other tests.

CE-6000 Specification
1Model
1. Material code CE-6008n-20V50A-H.pdf (61 downloads )
2Channels information
1. Channels quantity Channels quantity in one unit 8
2. Main channel Channel feature Constant current source and constant voltage source dual closed loop control
  Channel control mode Independent control
  Channel parallel connection Support max 4 channels parallel mode. Pulse and SIM tests will be disabled in channels parallel mode.

 

3Power grid side parameters
1.Input power   AC380V±15%  50/60±5Hz
2.Power factor   ≥99%(Full load)
3.THDi   ≤5%(Full load)
4.Input resistance   ≥1MΩ
5.Input power   9.4KW
6.Input current   14.3A/single
7.Overall system efficiency(Max)   90%
8.Noise   ≤65dB
9.Voltage and current sampling   Four-wire connection(same port for charging and discharging)
10.Power control module type   MOSFET
11.Input power wiring method   Three-phase-five wire system
12.Power input protection   Anti-surge, anti-silos, anti over or under frequency, anti over or under voltage, anti phase absence, etc.
4Functions and performances
1. Voltage Output range Charge:0V~20V
    Discharge:3V~20V
  Min discharge voltage 3V
  Accuracy ±0.1% of FS
  Resolution 24bit
2. Current Output range 0.25A~50A
  Accuracy(independent range) ±0.05% of FS
  CV cut-off current 50mA
  Resolution 24bit
3. Power Single channel output power 1KW
  Whole machine output power 8KW
4. Time Current response time ≤3ms
  Current conversion time ≤6ms
  Min. step time 0.1s
5. Charge/Discharge modes Charge/Discharge modes CCC, CVC, CC-CVC, CPC
    CCD, CVD, CPD, CRD
  Cut-off condition Voltage, Current, ΔTime, Capacity, -ΔV
6. Simulation Charge Current, Power
  Discharge Current, Power
  Switch Support continuous switching between charge and discharge
  Cut-off condition Time, step line
  Steps file lines 1,000,000
7. Pulse Mode Charge Current ,power
  Discharge Current, Power
  Min pulse 100ms
  Pulse counts Up to 32
  Charge and discharge switch supported
  Cut-off condition Voltage, ΔTime
8. DCIR   DCIR by calculation
9. Safely protection Software protection Power off data protection
    Offline mode function
    Safety protection conditions can be set, including:voltage lower limit ,voltage upper limit ,current lower limit ,current upper limit ,delay time, etc.
  Hardware protection Anti-reverse connection, over-voltage, over-current, over-temperature, etc.
5Data management and analysis
1. Step setting method   Form editing
2. Data report Recording conditions Minimum time interval: 10ms(connected with AUX channel:100ms)
    Minimum voltage interval: 0.04V
    Minimum current interval: 0.1A
  Recording frequency 100Hz(connected with AUX channel:10Hz)
3. Database   MySQL database
4. Data output   Excel, Txt
5. Curve type   Templates available, customization supported
6. Bar code scanning   Support bar-code scanning function
    Management and traceability of historical data
6Communication
1. Host computer communication   TCP/IP protocol
2. Communication port   Ethernet
3. Communication baud rate of the testers   1M
4. Host computer communication baud rate   10M~100M adaptive
5. Communication setup   Set up a LAN(local area network) through switches and routers
6. Communication expansion(optional)   Support CAN, RS485 communication and BMS communication, with DBC configuration function
7Environmental requirements, dimension and weight
1. Operation environment temperature   -10℃~40℃(When the temperature is 25±10℃, the accuracy error caused by temperature change is less than 0.005% of FS per degree)
2. Storage environment temperature   -20℃~50℃
3. Operation environment humidity   ≤70% RH(no moisture condensation)
4. Storage environment humidity   ≤80% RH(no moisture condensation)
5. Dimension W*D*H   600*800*1300(mm)
6. Weight   about 171.5KG

 

Weight 200 kg
Dimensions 65 × 85 × 135 cm
Brand

Neware

Model

CE-6008n-20V50A-H

Origin

China

Certificate

CE CNAS